Industrial electric conductivity meter (Two-Wire Transmitter)
Measuring range | Cell constant | /cm | 0.01 | 0.1 | 1.0 |
---|---|---|---|---|---|
Electrical conductivity | μS/cm | 0.000 to 2.000 0.00 to 10.00 | 0.000 to 2.000 0.00 to 20.00 0.0 to 100.0 | 0.0 to 200.0 0 to 1,000 | |
mS/m | 0.0000 to 0.2000 0.000 to 1.000 | 0.0000 to 0.2000 0.000 to 2.000 0.00 to 10.00 | 0.00 to 20.00 0.0 to 100.0 | ||
TDS conversion | mg/L | 0.00 to 2.00 0.0 to 10.0 | 0.0 to 20.0 0 to 100 | 0 to 200 0 to 1,0000 | |
Electrical conductivity measurement and TDS conversion measurement cannot be selected at same time. | |||||
Temperature | ËšC | 0 to 100 Resolution: 0.01ËšC | |||
Repeatability | Electrical conductivity: ±0.5% full-scale or less TDS conversion: ±1.5% full-scale or less Temperature: ±0.1˚C (for equivalent input) | ||||
Linearity | Electrical conductivity: ±0.5% full-scale or less TDS conversion: ±1.5% full-scale or less Temperature: ±0.5˚C (for equivalent input) | ||||
Transmission output | 4 to 20 mA DC Input/output insulated type Maximum load resistance 600 Ω | ||||
Contact input | One point Contact format: Non-voltage a contact Contact function: Transmission output is held for closed contact input. | ||||
Temperature compensation range | 0 to 100ËšC (but for 0ËšC or less, 100ËšC or more, extend and perform calculation) | ||||
Ambient temperature | -20 to 60ËšC | ||||
Temperature compensation element | Pt 1000 (0ËšC) | ||||
Calibration function | Electrical conductivity: Based on input of cell constant correction coefficient (parameter input) TDS conversion: Conversion based on specified coefficient (0.30 to 1.00) Temperature: One point calibration compared with reference temperature gauge | ||||
Self-diagnosis function | Sensor diagnosis error, Transmitter malfunction | ||||
Power supply | 24 V DC (operational voltage range: 21 V to 32 V DC) Power consumption 0.6 W or less | ||||
Construction | Outdoor installation type: IP65 protection level Mounting method: 50A pole or wall mounted Case: Aluminum alloy Mount fitting/hood: SUS304 | ||||
Compatible sensor | ESH series electrical conductivity sensor (cell constant 0.01/cm, 0.1/cm, 1.0/cm) | ||||
Weight | Approx. 4 kg | ||||
Equipment protected | CE marking, FCC rules |
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
For the C-131
For the C-121 and C-122
For Instant Water Quality Reports
Sanitary Conductivity Sensor (Ferrule Clamp type) for Low Concentrations
Sanitary Conductivity Sensor (Flow Through type) for Low Concentrations
OEM gratings design and production capabilities
Carbon Sensor Conductivity Meter (Low concentration type)
Industrial resistivity meter
KOH Monitor
Wide range TMAH Concentration Monitor
Sanitary Conductivity Sensor (Ferrule Clamp type) for Low Concentrations
Sanitary Conductivity Sensor (Flow Through type) for Low Concentrations
Carbon Sensor Conductivity Meter (Low concentration type)
Industrial resistivity meter
KOH Monitor
Wide range TMAH Concentration Monitor
High Precision TMAH Concentration Monitor
Carbon Sensor Resistivity Meter
2-Channel Resistivity Meter
Dissolved Oxygen Monitor in Low Concentration